Medical device inspection is a required step in both decontamination and prep & pack. Along with magnification and the recommended borescope inspection, there is one other important resource: Lighting. Often placed in areas of the hospital with minimal or poor access to natural light, sterile processing departments often feel as though they’re poorly lit, and they often are. ANSI/AAMI has provided guidance on the measure and quality of lighting that sterile processing and endoscopy departments should have.
Program Overview
Accredited by HSPA. Certificate issued automatically upon a passing score of 80/100 or higher.
Program Material
Medical device inspection is a required step in both decontamination and prep & pack. Along with magnification and the recommended borescope inspection, there is one other important resource: Lighting. Often placed in areas of the hospital with minimal or poor access to natural light, sterile processing departments often feel as though they’re poorly lit, and they often are. ANSI/AAMI has provided guidance on the measure and quality of lighting that sterile processing and endoscopy departments should have.
ANSI/AAMI ST79 and ST91 both detail guidelines for lighting, as well definitive illuminance for given work areas and functions, according to the recommendations by the Illuminating Engineering Society of North America (IES).
| Work area/function | Least Illuminance | Average Illuminance | Highest Illuminance |
| 500 lux
(50 foot-candles) |
750 lux
(75 foot-candles) |
1000 lux
(100 foot-candles) |
|
| Detailed Inspection | 1000 lux
(100 foot-candles) |
1500 lux
(150 foot-candles) |
2000 lux
(200 foot-candles) |
| Sink Areas | 500 lux
(50 foot-candles) |
750 lux
(75 foot-candles) |
1000 lux
(100 foot-candles) |
| General Work Areas | 200 lux
(20 foot-candles) |
300 lux
(30 foot-candles) |
500 lux
(50 foot-candles) |
| Processed Storage | 200 lux
(20 foot-candles) |
300 lux
(30 foot-candles) |
500 lux
(50 foot-candles) |
*See table in ANSI/AAMI ST79:2017 (4.3.8 Lighting) and ANSI/AAMI ST91:2021 (3.3.5.6 Lighting).
These lighting recommendations are not one-size-fits-all. Three considerations need to also be accounted for when assessing the optimal illumination of a working area in a reprocessing department:
The location within the facility isn’t the only hinderance when it comes to lighting quality. Many factors can contribute to poorly lit sterile processing or gastroenterology departments:

Lighting is a concern many SPD and GI departments struggle with, but it doesn’t have to be. ANSI/AAMI standards point to specific luminosities for various working areas, with consideration for employees’ ages, nature of work, and reflectivity of the working environment in mind. Addressing some common challenges, such as removing obstructions, implementing adjustable lighting options, and lessening the reliance on overhead lighting, are easy ways departments can start improving their lighting conditions.
Interactive Learning
Get #1 on the free play scoreboard by the end of SP Week (10/18), and win an extremely rare, exclusive Pure E.D.U. pin. Welcome to Light It Right, an inspection game where you choose the right light, set it until your view is ready, and sweep each instrument for residue, damage, leaks, and holes in wrap. Education Mode (Training challenges) teaches one lighting idea per scene, with a quick learning check after each, and doesn’t count toward the scoreboard. Free Play (Inspection shift) counts toward the scoreboard: random findings, three minutes per scene, and five lives. When your shift ends, enter your first name and email to save your score. Only your first name is shown, and your email is used just to contact the winner. Play on your computer or phone, and tap Full screen for the best view. The game is for fun and review and doesn’t earn CE credit, so take the quiz below to claim your 0.5 CE.